An article explaining the working principle and maintenance of sound level meters

Jan 20, 2024

Leave a message

An article explaining the working principle and maintenance of sound level meters

 

working principle
The microphone converts the sound into an electrical signal, and then the preamplifier changes the impedance to match the microphone and the attenuator. The amplifier adds the output signal to the weighting network, performs frequency weighting on the signal (or external filter), and then amplifies the signal to a certain amplitude through the attenuator and amplifier, and sends it to the effective value detector (or external voltage detector). Level recorder), giving the value of the noise level on the indicator head.


microphone
A microphone is a device that converts sound pressure signals into voltage signals. It is also called a microphone. It is the sensor of a sound level meter. Common microphones include crystal, electret, moving coil and condenser.


1. The dynamic microphone is composed of a vibrating diaphragm, a movable coil, a magnet and a transformer.


sound level meter
The vibrating diaphragm begins to vibrate after being subjected to sound wave pressure, and drives the movable coil installed with it to vibrate in the magnetic field to generate induced current. This current changes according to the amount of sound wave pressure the vibrating diaphragm is subjected to. The greater the sound pressure, the greater the current generated; the lower the sound pressure, the smaller the current generated.


2. The condenser microphone is mainly composed of a metal diaphragm and closely spaced metal electrodes. It is essentially a flat capacitor. The metal diaphragm and the metal electrode constitute the two plates of the flat capacitor. When the diaphragm is affected by sound pressure, the diaphragm deforms, causing the distance between the two plates to change, thus changing the capacitance. The voltage in the bit measurement circuit also changes, realizing the function of converting the sound pressure signal into a voltage signal. Condenser microphone is an ideal microphone in acoustic measurement. It has the advantages of large dynamic range, flat frequency response, high sensitivity and good stability in general measurement environments, so it is widely used. Since the output impedance of the condenser microphone is very high, impedance conversion needs to be performed through a preamplifier. The preamplifier is installed inside the sound level meter close to where the condenser microphone is installed.


amplifier
Generally, two-stage amplifiers are used, namely input amplifier and output amplifier, whose function is to amplify weak electrical signals. The input attenuator and output attenuator are used to change the attenuation of the input signal and the attenuation of the output signal so that the meter pointer points at the appropriate position. The attenuator adjustment range used by the input amplifier is the measurement low end, and the attenuator adjustment range used by the output amplifier is the measurement range. Many sound level meters have a cutoff of 70dB for the high and low ends.


weighted network
In order to simulate the different sensitivities of human hearing at different frequencies, the sound level meter is equipped with a network that can simulate the hearing characteristics of the human ear and modify the electrical signal to an approximate value of the sense of hearing. This network is called a weighting network . The sound pressure level measured through the weighting network is no longer the sound pressure level of an objective physical quantity (called linear sound pressure level), but the sound pressure level modified by the sense of hearing, called weighted sound level or noise level.


Maintenance
⑴Keep the outside of the instrument clean;


⑵Microphones should be stored dry when not in use;


⑶The microphone diaphragm should be kept clean and should not be touched with hands;


⑷ When the instrument is not used for a long time, it should be powered on for 2 hours every month and 2 hours a week during the mildew season;


⑸The battery should be removed in time after using the instrument;


⑹Send to the measurement department for verification regularly;

 