Common problems and solutions with the microscope

Apr 19, 2024

Leave a message

Common problems and solutions with the microscope

 

First, the microscope bulb does not light

1, check whether the power cord is loose, whether the socket is normal.

2, check whether the fuse is blown (fuse more in the body of the power connector), such as blown fuse replacement.

3, replace the spare bulb, pay attention to the replacement process, do not directly touch the fingers to the bulb.

 

Second, the microscope can not focus imaging

1, check whether the focus limit mechanism is not adjusted to cause the working distance is not enough.

2, check whether the sample is put in reverse, if the coverslip down in the high power working distance is not enough.

 

Third, the image is clear under low magnification, high magnification imaging blurred

1, Higher magnification of the sample thickness and flatness have higher requirements.

2, biological 100x lens should pay attention to whether there is no oil dripping, whether the objective lens is clean.

3, switch to high magnification when the aperture diaphragm on the condenser mirror to play big accordingly.

 

Fourth, the field of view appears bright uneven

Check whether the optical path is centred before using the microscope.

 

Fifth, the light bulb has been lit, but the eyepiece field of view to observe all black

1, check whether the objective lens into the light path, field of view diaphragm and aperture diaphragm is adjusted too small, whether the light path is centred.

2, check whether the trinocular head eyepiece and video switching lever hit in the video position.

 

Sixth, the brightness of the field of view is dark

1, the field of view diaphragm and aperture diaphragm is adjusted too small.

2, whether the optical path is adjusted in place (condenser, diaphragm, filament).

3, whether the use of polarisation, phase difference components.

 

Seven, the eyepiece field of view of black edges

1, check whether the field of view diaphragm playing too small;

2, the optical path is not centred (spotting scope, diaphragm, filament);

3, the objective lens carousel whether to play in place;

4, the eyepiece barrel and the body of the card position is accurate.

 

Eight, synchronisation problems, the use of camera observation, eyepiece observation is clear, but the computer shows a blurred image

1, adjust the interface or interface lens height to achieve synchronisation.

2, adjust the interface can not be fully synchronised, try to adjust the eyepiece diopter.

 