Current Test Accessories for Multimeters
Sometimes the current to be measured exceeds the range of the digital multimeter, or on-site conditions do not allow opening the circuit for current measurement. For high-current tests (typically above 2 A) where high accuracy is not required, a current clamp is very convenient. The current clamp clamps around the conductor to measure the current and converts it into a signal readable by the digital multimeter.
There are two basic types of current probes: current transformer types, for AC measurement only; and Hall‑effect probes, for measuring AC or DC current.
Current transformer type It generally outputs 1 mA per 1 A of measured current. A 100 A current is converted to 100 mA, which can be easily measured by a digital multimeter. Connect the test leads to the mA and COM jacks, and turn the function selector switch to the AC milliamps range.
Hall‑effect probe It outputs 1 mV per 1 A of AC or DC current. For example, an AC current of 100 A is converted to an AC voltage of 100 mV. Connect the test leads to the V and COM jacks, and set the function selector to the V or mV range. The multimeter now displays 1 mV to represent 1 A of current.
Always cut off power to the circuit under test before opening or unsoldering the circuit for current measurement. Even small currents can be hazardous.
Do not measure voltage while test leads are plugged into the current jacks. This may damage the meter and create an electric shock hazard.
Transformer‑style current probes, such as the Fluke 80i-400, use a 1 mA reading on the digital multimeter to represent an actual 1 A signal.
The Fluke I-1010 Hall‑effect probe can measure very high AC or DC currents. It converts current into a voltage signal. A voltage reading of 1 mV represents an actual current of 1 A.