Detector Characteristics of Noise Meter

Feb 13, 2023

Detector Characteristics of Noise Meter

 

For dosimeters with "calibration" gears, the calibration method can refer to the verification regulations. For most dosimeters without "calibration" gears, follow the steps below:


a. Set the amplitude of the 1000Hz sinusoidal signal applied to the dosimeter according to the standard system indicated sound level in Table 1 of this article;


b. Apply the signal from the previous step to the dosimeter, and record the duration when the dosimeter shows 100%;


c. Add the sound burst with a frequency of 2000Hz to the standard system, and adjust the signal amplitude so that it is the same as that of step (a);


d. Apply the sound burst from the previous step to the dosimeter, and record the displayed value of the dosimeter after the same duration as in step (b). For Class 2 dosimeters, the indicated value should be between 79% and 126%.

 