How to change the electrolyte of the micro moisture meter

 

Trace moisture tester is a commonly used instrument in the laboratory, in its manual, has been comprehensively introduced to the use of the instrument, assembly and maintenance operations, but these are basic operations, in ordinary use as long as attention is paid to the maintenance, this problem can be avoided. In order to ensure the accuracy of the trace moisture detector determination, usually pay attention to the electrolyte replacement.


Before replacing the electrolyte, prepare a clean filter paper, and then put the two dry tubes in the electrolytic cell on the paper, and then take out the electrolytic electrode, pour out the electrolyte inside, and then pour out the electrolyte in the electrolytic cell slowly, it is worth noting that: in the process of pouring the electrolyte, don't pour out the stirrer.


In general, there is no need to clean the electrolytic cell, but if it is really polluted, it can be cleaned, but it should be dried after cleaning. During the cleaning process, pay special attention to the white plate at the bottom of the electrolysis electrode, because it is a kind of honeycomb ceramic plate, once it contains water, it is difficult to be released, so as far as possible, do not use water for cleaning.


There is really no way, have to use water to clean the time, after cleaning should be placed in the temperature of about 50 ℃ oven drying, time can be controlled in 6 hours or so, and then let the electrolytic electrode bottom of the trace moisture detector of the white plate, in the oven naturally for cooling.

 