1) Carefully read the instruction manual and follow the guidelines it lays out before beginning the process of daily use and verification.
(2) Correctly change the gas detection device's time and date to make it easier to use the device for the subsequent instrument verification.
(3) A stable voltage and enough power must be used with the device.
(4) Avoid having water or other liquids sucked into the detector when performing pump sampling.
(5) To zero-calibrate the instrument, use pure air or nitrogen.
(6) Use the same standard gas as the observed gas to calibrate sensitivity, with the standard gas concentration being 60% of the detection instrument's full scale.
(7) To confirm the alarm concentration set point of the instrument, use a gas standard material that is 1.5 times the alarm concentration set value.
(8) The gas used in the verification must be either the standard gas generated by the standard gas distribution device certified by metrology, or the standard gas approved and promulgated by the metrology administrative department and provided by the unit with the corresponding standard material "License for Manufacture of Measuring Instruments."
(9) The apparatus used for testing consists of the following: gas standard material, flow controller, zero-point gas, stopwatch, decompression valve and gas circuit, calibration cover, insulation resistance meter, and stabilized power supply.
(10) The instrument should be calibrated on a regular basis, and the calibration interval shouldn't be longer than a year.
The objective requirement of contemporary industry is safety; in order to produce, production must be safe. Gas detection tools and meters are receiving more and more attention as a crucial component of ensuring safe production. Additionally, the number of companies making gas detection tools is increasing. Don't forget to select the most appropriate option—not the most expensive. In order to protect the safety of both production and life, you should make sure the instrument is functioning normally and is in good condition while in use.