Methods for Storage and Maintenance of Electrodes of pH Meters (Acid Meters)

Oct 26, 2023

Leave a message

Methods for Storage and Maintenance of Electrodes of pH Meters (Acid Meters)

 

pH meter (acidity meter) electrode storage and maintenance methods


How to store pH meter (acidity meter) electrodes:
Please store the electrode in 3mol/L potassium chloride solution (3M KCl) when not in use; it can be stored in pH7.00 buffer for a short period of time. Leaving the pH meter (acidity meter) electrode dry or soaked in distilled water for a long time will shorten the service life of the electrode.


pH meter (acidity meter) electrode maintenance methods:
After using the pH meter (acidity meter) electrode for a period of time, if you find that the slope becomes lower and the response speed becomes slower, you can try the following methods.
1. If the measurement sample contains protein, pepsin/hydrochloric acid washing solution (ME51340068) can be used to clean the electrode membrane.
2. If the measurement sample is oily/organic liquid, it can be rinsed with acetone or ethanol.
3. If the junction of the pH meter (acidity meter) electrode is found to be dirty and black, you can use mercaptan cleaning solution (ME51340070) to clean the junction.
4. Activation of electrode membrane.


Activation method: Soak the pH meter (acidity meter) electrode regeneration solution (ME51340073) for 30 seconds, then soak in 3mol/L KCl solution for 5 hours


Display content of explosion-proof industrial online pH meter in measurement operation mode
1. Measured value and time (24/12 h AM/PM) and unit
Is the temperature in °C or °F (format selectable in configuration)
Press the meas button in the measurement mode, and the following display screen appears (lasts about 60 seconds):


2. Selection of measured values and parameter group A/B (if configured)


3. Measurement value and measurement point name ("TAG", the measurement point name can also be entered in the cultivation)


4. Time and date


Note: Pressing the enter button on Mettler's explosion-proof industrial online pH meter will display the output current. The display appears as long as the enter button is pressed and then switches back to the measured value display after 3 seconds.

 

2 water ph meters

Send Inquiry