+86-18822802390

Regulations to be observed when using a coating thickness gauge

Nov 22, 2022

Regulations to be observed when using a coating thickness gauge


a Base metal properties


For the magnetic method, the magnetic properties and surface roughness of the base metal of the standard piece shall be similar to those of the test piece base metal. For the eddy current method, the electrical properties of the base metal of the standard sheet shall be similar to those of the test piece base metal.


b base metal thickness


Check whether the base metal thickness exceeds the critical thickness, if not, use one of the methods in 3.3 to calibrate.


c edge effect


Measurements should not be made close to sudden changes in the test piece, such as edges, holes and inner corners.


d Curvature


Measurements shall not be made on curved surfaces of test pieces.


e Number of readings


Often several readings must be taken within each measurement area since each reading from the instrument is not identical. Local variations in cover thickness also require multiple measurements in any given area, especially when the surface is rough.


f surface cleanliness


Before measurement, any attached substances on the surface should be removed, such as dust, grease and corrosion products, etc., but do not remove any covering substances


The difference between F, N and FN in the coating thickness gauge:


F stands for ferrous ferromagnetic substrate. The F-type coating thickness gauge adopts the principle of electromagnetic induction to measure non-ferromagnetic coatings and coatings on ferromagnetic metal substrates such as steel and iron, such as: paint, powder, plastic, rubber , synthetic materials, phosphating layer, chromium, zinc, lead, aluminum, tin, cadmium, porcelain, enamel, oxide layer, etc.


N stands for Non-ferrous non-ferrous substrate, and the N-type coating thickness gauge adopts the principle of eddy current; to measure the enamel, rubber, paint, plastic layer, etc. on copper, aluminum, zinc, tin and other substrates with eddy current sensors.


The FN type coating thickness gauge adopts both the principle of electromagnetic induction and the principle of eddy current. It is a two-in-one coating thickness gauge of F type and N type.


AR932--3