Standard Preparation Method of pH Meter Calibration Solutions

Jun 26, 2026

Standard Preparation Method of pH Meter Calibration Solutions

 

The pH meter should be regularly calibrated to ensure that its precision and accuracy meet the requirements. How to calibrate a pH meter, or in other words, how to calibrate an acidity meter. The following is the information compiled by the author. Let's take a look at the calibration solution and method of the pH meter and learn about them together.
Standard method for preparing pH meter calibration solution
1. PH4 potassium hydrogen phthalate standard buffer solution:
Accurately weigh 10.12g of potassium hydrogen phthalate [KHC8H4O4] dried at 115 ± 5 ℃ for 2-3 hours, dissolve and dilute to 1000ml with water.
2. PH7 phosphate standard buffer solution (pH7.4:
Accurately weigh 4.303g of anhydrous disodium hydrogen phosphate and 1.179g of potassium dihydrogen phosphate dried at 115 ± 5 ℃ for 2-

 

3 hours, dissolve and dilute to 1000ml with water.
Additional information: Phosphate standard buffer solution (pH 6.8), accurately weigh 3.533g of anhydrous disodium hydrogen phosphate and 3.387g of potassium dihydrogen phosphate dried at 115 ± 5 ℃ for 2-3 hours, add water to dissolve and dilute to 1000ml.

 

3. PH9 borax standard buffer solution:
Accurately weigh 3.80g of borax [Na2B4O7 · 10H2O] (note: avoid weathering, dissolve and dilute to 1000ml with water, place in a polyethylene plastic bottle, seal tightly, and avoid contact with carbon dioxide in the air).

 

Working principle of pH meter
A pH meter is used to measure the pH value of a solution using the potentiometric method. Therefore, the working mode of a pH meter is not only to measure the pH value of the solution, but also to measure the electromotive force of the battery. PH, in Latin, is the abbreviation for Pondus hydrogeni, which is the activity of hydrogen ions in a substance. PH is the negative logarithm of the concentration of hydrogen ions.

 

The main measuring components of a pH meter are a glass electrode and a reference electrode. The glass electrode is sensitive to pH, while the potential of the reference electrode is stable. Putting the two electrodes of the pH meter together into the same solution forms a primary cell, and the potential of this primary cell is the algebraic sum of the potentials of the glass electrode and the reference electrode.

 

The reference electrode potential of a pH meter is stable, so under stable temperature conditions, the potential change of the original battery composed of solution and electrode is only related to the potential of the glass electrode, which depends on the pH value of the solution being measured. Therefore, by measuring the potential change, the pH value of the pH solution can be obtained.

 