Thickness Gauge Coupler Instructions
1. Thickness gauge need to cooperate with the coupling agent to be normal, the purpose of the coupling agent is to exclude the probe and the measured object before the air, and through the coupling agent on the ultrasonic signal conduction.
2. Use the amount of the size of a grain of rice can be, after the ultrasonic probe can be attached.
3. Ultrasonic coupling agent with rust, corrosion, antioxidant properties, if the use of exhaustion is recommended to buy from regular manufacturers or regular agents.
4. Emergency measures: coupling agent run out, temporarily can not buy the case, you can temporarily use viscous liquid (such as hand soap, laundry detergent, medical ultrasound oil) to replace, but the above alternatives do not have antioxidant properties and signal conductivity is poor, only a temporary alternative to temporary use, after use, please be sure to use dry paper towels to ultrasonic probe and the material of the liquid on the wipe clean.
27MG Thickness Gauge calibration steps before use
The 27MG thickness gauge must be calibrated before use to ensure that it gives accurate readings when the probe is used to make measurements. The calibration process involves calibrating the speed of sound and the zero position on two known thicknesses of a test block (e.g., a fifth-order block, as shown in the figure below) of the exact same material as the block being tested.
Calibration steps:
1, on the surface of the test block thick steps drop coupling agent
2, will be coupled to the probe in the test block of thick steps
3, press 【CAL VEL】or 【VAL v】(calibrate the speed of sound key)
4, the thickness reading of the display stabilised by [ENTER] key or [√] (OK key)
6, press [CAL ZERO] or [CAL Ø] (calibration zero key)
7, in the test block thin step of the surface drops of coupling agent
8, the probe coupled to the thin steps of the test block
9, the thickness reading of the display stabilised by [ENTER] key or [√] (OK key)
10, use the arrow keys to enter the thickness reading
11, press 【MEAS】(Measurement) key