Trace Moisture Meter Frequently Asked Questions

Mar 20, 2024

Trace Moisture Meter Frequently Asked Questions

 

1, the anode electrolyte colour is not bright yellow, but between brown and dark yellow.
Reason: moisture meter electrolyte colour is too dark, is the electrode on the electrolyte response ability to reduce.
Solution: you can use paper towels to clean the double platinum needle electrode to remove the surface of the adsorbent; check whether the measurement electrode is normally connected; the measurement electrode may be malfunctioning.


2,Pre-titration of fresh anode electrolyte, the drift is too high
Reason: There is residual water in the titration system.
Solution: you can replace the molecular sieve and silica gel in the drying tube, check whether the electrode interfaces and plug interfaces of the titration table are tight; you can apply silicone grease on some loose interfaces.

3,Why the drift is too high during standby titration?
Reason: The water in the cathode pool penetrates into the anode pool through the diaphragm.
Solution 1: Replace the electrolyte in the anode cell; add a small amount of one-component volumetric Karl Fischer reagent to the cathode cell for drying; keep the liquid level of the anode liquid higher than the liquid level in the cathode cell.


Solution 2: Clean the titration cup effectively to remove the continuous side reaction caused by the sample left over from the last test; check the sealing of the titration system.


4,High drift value after sample titration
Reason: The test sample has a side reaction with the anode electrolyte and the reaction produces water.
Solution 1: Replace other kinds of anode electrolyte or replace other sample pretreatment methods


Solution 2: Some volatile components in the sample are allowed to react with the Karl Fischer reagent.


Solution 3: Improve the sample pretreatment method.


5,Long titration time, titration does not stop
Reason 1: Improper selection of control parameters.
Solution: Relative drift termination can be used as the end parameter, increase the relative drift termination value, increase the end point.


Reason 2: Anode electrolyte conductivity is too low.


Solution: Need to replace the anode electrolyte.

 