Treatment method for scaling of PH meter probe
For general scaling use a non-abrasive household cleaner
Scale or metal hydroxide use diluted hydrochloric acid about 0.1%-3% for 1-5 minutes
Oily lubricants use solvents such as alcohol or acetone
Biological contaminants use a diluted hydrochloric acid and pepsin mixture for several hours
Do not use solvents such as acetone to clean the electrodes as they can damage the plastic rod
The electrode probe of the industrial online ph meter must be rinsed thoroughly after cleaning
In addition, the metal surface of the industrial online redox electrode can be ground and polished with sandpaper
If the ceramic diaphragm of the reference electrode installed later is blocked, it can be cleaned like the glass electrode of the industrial online pH meter. In addition, it can be carefully scraped off, but care must be taken not to scratch the diaphragm