Troubleshooting Microscope Mechanisms
1, rough adjustment part of the troubleshooting
Coarse adjustment of the main fault is the automatic decline or lifting and lowering of the tension is not the same. The so-called automatic decline refers to the mirror tube, mirror arm or carrier stationary in a certain position, without adjustment, under the action of its own weight, the phenomenon of falling down automatically and slowly. The reason is the mirror tube, mirror arm, the carrier itself gravity is greater than the static friction caused. The solution is to increase the static friction, so that it is greater than the gravity of the mirror tube or mirror arm itself.
For oblique cylinder and most of the binocular microscope coarse adjustment mechanism, when the mirror arm automatically slides down, can be used in both hands to coarse adjustment of the handwheel held to the inside of the stop pulley, both hands are tightened in the clockwise direction, you can stop the slide. If it does not work, you should find a professional to repair.
Mirror tube automatically down, often give people the illusion that the gear and rack with too loose caused. So add shims under the rack. In this way, although the slide of the mirror tube can be temporarily stopped, but it makes the gear and rack in an abnormal bite state. As a result of the movement, the gear and rack are deformed. Especially when the pad is not flat, the deformation of the rack is more powerful, resulting in part of the bite is tight, part of the bite is loose. Therefore, this method should not be used.
In addition, due to the rough adjustment mechanism for a long time in disrepair, lubricant dry, lift will produce uncomfortable feeling, and even can hear the friction sound of the machine parts. At this time, the mechanism can be removed and cleaned, reassembled after greasing.
2, fine-tuning part of the troubleshooting
Fine-tuning part of the most common failure is stuck and failure. The fine-tuning part is installed inside the instrument, and its mechanical parts are small and compact, which is the most delicate and complex part of the microscope. Failure of the fine adjustment part should be repaired by professional technicians. Do not dismantle it without sufficient certainty.
3, the objective lens converter troubleshooting
The main failure of the objective lens converter is the positioning device failure. Generally positioning spring plate damage (deformation, fracture, loss of elasticity, spring plate fixing screws loose, etc.) caused. Replacement of the new spring plate, do not tighten the set screw, should be in accordance with this section "three (b) 2" for the first optical axis correction. After the axis is merged, then tighten the screws. If it is the internal positioning type converter, you should screw down the big head screw in the centre of the rotating disc, remove the rotating disc, in order to replace the positioning of the spring plate, the optical axis correction method is the same as the previous.