Universal Tool Microscope Three Precautions in Operation
1, pay attention to the eyepiece and objective lens focusing order
Many people start measuring the focus with the objective lens, when the object focus is adjusted and then use the eyepiece in the "metre" line to align the measurement, if the "metre" line is not clear enough, the eyepiece will be focused. In fact, this order is wrong, because it will cause the image of the measured object in front of the focus after the existence of a certain shadow. The correct method is the first eyepiece in the "metre" line clear, and then focus on the object, so as to ensure that the "metre" line and the image of the object is clear.
2, pay attention to the measurement of before the surface of the measured part of the burr and knock marks
The measured parts may produce some burrs and knocks in the process, use and transport, these defects are not easy to be detected, but in the measurement of the Universal Tool Microscope is easy to cause alignment errors or cause the measurement surface is not on the same focal plane and the formation of a certain local virtual shadow, thus affecting the accuracy of the measurement results, so be sure to thoroughly ** these surface burrs and knocks.
3,Pay attention to the correct installation of the measured parts
Universal tool microscope on the measured part of the installation form generally have two kinds.
(1) plane test piece placement. For plane test pieces of the main attention to be measured by the measured surface should be in the same focal plane, otherwise it is easy to form a local virtual shadow, for the measured surface chamfered parts, * good to let the chamfer face down, otherwise it is easy to cause focus is not clear, resulting in inaccurate measurement.
(2) Installation of shaft test pieces. Shaft measuring parts generally rely on the central hole positioning, before installation, be sure to clean the hole, in particular, to eliminate the sediment and burrs, otherwise it will cause the axis of the measured parts and the centre line of the instrument is not the same axis, which will bring a large measurement error. This situation is often encountered in the daily measurement, the best way is to use the instrument after the installation of the horizontal line of the "metre" line in the demarcation board to check the outside diameter of the measured shaft runout error, so as to determine whether the measured parts are installed correctly.