With a gas detector, what is the difference between testing and calibrating?

Mar 12, 2025

Leave a message

With a gas detector, what is the difference between testing and calibrating?

 

After a period of use, gas detectors are easily affected by the usage environment and the gas sensor of the instrument itself, which may result in significant deviations in the measured results. Therefore, in order to ensure the accuracy of measurement results, regardless of the type of gas sensor used in the gas detector, regular testing is necessary. If the deviation of the test results exceeds the normal range, the gas detector needs to be recalibrated. Many friends often confuse testing and calibration, so what is the difference between gas detector testing and calibration?


The difference between gas detector testing and calibration:
(1) Testing refers to using a gas detector to detect gases of known concentration, in order to determine whether the results detected by the instrument are within an acceptable range. If they exceed the allowable range, the instrument needs to be recalibrated.


(2) Calibration refers to adjusting the results of a gas detector with a known concentration of gas to match the concentration of the known gas.


Frequency of gas detector testing and calibration:
(1) If conditions permit, gas detectors should be tested once a day before use;


(2) For gas detectors that fail the test, they must be calibrated before use;


(3) If the tested environment may affect the performance of the gas detector, the test should be conducted at any time.


If the conditions do not allow for daily calibration confirmation, the gas detector can reduce the calibration frequency under the following conditions:
(1) At least 10 days of testing were conducted in specific occasions, and the daily test results confirmed that the gas detector was not affected by certain gases in the environment that caused gas sensor poisoning.


(2) If it is determined after testing that the gas detector does not require calibration, the calibration interval can be extended, but the duration cannot exceed 30 days.


(3) The calibration history of the instrument should be managed by a dedicated person or have detailed tracking and usage record data files.

 