6 Common Faults of Clamp Ammeter

May 26, 2023

Leave a message

6 Common Faults of Clamp Ammeter

 

(1) When measuring current or voltage, there is no indication for one or several gears, and the other gears are normal. The reason is that the fastening switch screw of the clamp meter is loose, or the connection on the branch switch is twisted. This kind of Faults are the most common, just connect the broken wire after opening the cover.


(2) The reading of the current gear is too small, and the reading of the voltage gear is normal. Most of the reasons are caused by poor contact of the jaws and too much magnetic flux leakage. The jaws should be corrected to make them in good contact. This kind of failure is sometimes caused by a short circuit between windings. It is more troublesome at this time. Generally speaking, it needs to be rewound according to the original data, and aging treatment is also required.


(3) The readings of the current and voltage files are all low. This phenomenon often occurs on the internal magnetic clamp meter. The reason is that the magnet is demagnetized. Generally, it should be magnetized to solve it. It can also be solved by reducing the resistance connected in series with the branch circuit of the meter head. value is adjusted.


(4) One of the readings is inaccurate, and the corresponding resistance should be adjusted.


(5) The indication of the voltage level is normal, but there is no indication of all the current levels. Use a multimeter to check the primary wire switch and secondary winding to see if there is any abnormality.


(6) There are no indications at all. Check the rectifier diode, meter head, switch, and the connection with the clamp meter or related wiring for breaks.

 