The sugar meter is used to swiftly ascertain the refractive index or concentration of liquids that include sugar as well as other non-sugar solutions. widely utilized in agricultural production, scientific research, the food, beverage, and other industrial sectors (including sugar). It is appropriate for measuring the concentration of different sauce (seasoning) items such soy sauce and tomato sauce as well as the amount of sugar in jam, syrup, liquid sugar, and other products with a high sugar content.It is appropriate for pre-delivery inspection, quality management, the production line for carbonated drinks, refreshing drinks, and fruit juices, among other things. It works well for the entire fruit production cycle, from planting to selling. It can be used to classify the sweetness and pinpoint the precise harvest time. Additionally, the textile sector makes extensive use of the assessment of pulp concentration.
Calibration and temperature correction of the sugar meter:
Prior to measurement, the sugar meter needs to be zero-calibrated. In order to set the dividing line to the 0% point of the scale, take a few drops of distilled water, place them on the detecting prism, and turn the zero adjustment screw. Then remove the detection prism by wiping it. When calibrating, some instrument models require standard solution in place of distilled water.
Use the temperature correction table to add (or remove) the temperature correction value from the value obtained at ambient temperature in order to obtain an accurate reading. This approach is solely appropriate for the assessment of sugar content.