Emergency In-Circuit Resistance Measurement with a Digital Multimeter

Nov 21, 2026

Emergency In-Circuit Resistance Measurement with a Digital Multimeter

 

(1) Different models of digital multimeters have different test voltages and open‑circuit voltages for the resistance range. Therefore, the value of the loading resistor R1 shall be determined by experiment.

 

(2) First connect the loading resistor R1 between the V/Ω and COM jacks of the digital multimeter, and read the measured value of R1 with the multimeter before carrying out in‑circuit resistance measurement. Do not connect R1 in parallel to the circuit under test directly. If you do so, the relatively high test voltage of the multimeter resistance range may turn on transistors in the circuit and render the measurement meaningless (introduce large measurement errors). This test sequence must be strictly followed.

 

(3) In typical circuits, resistors connected in parallel with the emitter and collector of transistors are usually in the range of several kΩ to hundreds of kΩ, while resistors of several tens of ohms are less common. For this reason, the 200 kΩ range (with a resolution of 0.1 kΩ) is generally selected for measurement. If the reading shows R = R1 × RX / (R1 + RX) ≈ 0, it indicates a short‑circuit fault in the circuit under test (RX ≈ 0). Switch to the low 2 kΩ range for further inspection. If the reading R = R1 × RX / (R1 + RX) is close to R1, an open‑circuit fault may exist in the circuit under test (RX ≈ ∞), or the selected range is too low. Switch to the higher 2 MΩ range for re‑measurement.

 

(4) The 200 Ω and 20 MΩ resistance ranges are seldom used for in‑circuit measurement. When the loading resistor R1 is connected in parallel with the resistor RX under test, the actual measurable resistance range is extended, and the capability to judge high‑resistance faults is improved. The 2 MΩ range is commonly adopted. Besides, the resolution of the 2 kΩ range is already sufficient for judging whether the linear transistor under test is intact. Generally, only three loading resistors are required for in‑circuit resistance measurement. Take the DT830A digital multimeter as an example:

2 kΩ range: R1 = 1 kΩ

200 kΩ range: R1 = 47 kΩ

2 MΩ range: R1 = 470 kΩ We may also replace the three resistors with a single 470 kΩ resistor.

 

(5) After finishing in‑circuit measurement, remember to remove the loading resistor R1 connected between the V/Ω and COM jacks of the multimeter. This avoids affecting normal use of the multimeter and prevents hazards during high‑voltage measurement.

 