Introduction of related calibration schemes for infrared thermometers
The non-contact infrared thermometer consists of three series of portable, online scanning, multiple options and computer software, each with a variety of models and specifications. Among thermometers of various specifications, it is very important for users to choose the correct infrared thermometer.
Calibration time and self-calibration procedure of infrared thermometer
1. Range: The self-calibration procedure for infrared thermometers is used for the calibration of infrared thermometers that measure functions within the range (0°C to 1600°C).
2. Purpose: To calibrate the temperature of the heating furnace, heat treatment furnace and other equipment used in the company's production process and the infrared thermometer required for the temperature control measurement of the workpiece product during the production process, so as to ensure that the results meet the measurement accuracy requirements.
3. References: Product Technical Specification "Verification Regulations for Working Total Radiation Thermometers"
4. Overview: The infrared thermometer is a non-contact measuring instrument, easy to use, and can quickly measure non-contact and non-destructive temperature.
5. Calibration procedure
To establish a baseline for calibration:
According to the temperature measurement range, put a metal material sample with a diameter area larger than the target diameter required by the infrared thermometer in the resistance furnace.
A certified S-type precious metal thermocouple is bound on it, and the thermocouple is used as a reference temperature source.
Ensure that the temperature of the temperature source is relatively stable, and the expansion uncertainty of the reference calibration temperature source should not exceed the table.
Calibration method: Calibration environment conditions Ambient temperature (18-25) ℃ Reference standard The ambient humidity of the electrical measuring device should meet the corresponding technical requirements.
Relative humidity: >85%, avoid strong background radiation and alternating magnetic field.
Supporting facilities: The accuracy (converted into temperature) of the S-type thermocouple used to measure the standard temperature source and its point test equipment should be better than one tenth of the maximum allowable error of the infrared measuring instrument.
Insulation resistance meter with a rated DC voltage of 520V, temperature source heating furnace calibration work frame (table) meter ruler.
When starting up, the digital display should be clear and complete, and the buttons should be normal. If there is a warning that the battery is low, the calibration can only be performed after replacing the battery.
6. Calibration result processing:
Calibration data processing is carried out in accordance with the verification regulations for working-use full radiation thermometers. During the calibration process, the infrared thermometer calibration record form is written, and the calibration report is filled out after the calibration is completed.
7. Calibration time interval Calibration time interval generally does not exceed 1 year.