Introduction to the Service Life of Portable Gas Detectors

Jan 14, 2026

Leave a message

Introduction to the Service Life of Portable Gas Detectors

 

A, During the use of portable gas detectors, it is important to pay attention to regular calibration and testing

At present, many gas detectors can replace their detection sensors, but this does not mean that a detector can be equipped with different detector probes at any time. Whenever replacing the probe, in addition to requiring a certain sensor activation time, the instrument must also be recalibrated. In addition, it is recommended to perform response testing on the standard gas used in various instruments before use to ensure that the instruments truly provide protection. Combustible gas detectors and toxic gas detectors, like other analytical and detection instruments, are measured using a relative comparison method: first, a zero gas and a standard concentration gas are used to calibrate the instrument, and the standard curve is stored in the instrument. When measuring, the instrument compares the electrical signal generated by the concentration of the gas to be measured with the electrical signal of the standard concentration to calculate the accurate gas concentration value. Therefore, zeroing the instrument at any time and regularly calibrating the instrument are essential tasks to ensure accurate measurement.

 

B, Pay attention to the detection interference between different sensors

Although a composite gas detector can simultaneously detect multiple other gases, a portable gas detector is not an arbitrary combination of toxic and harmful gases. Generally speaking, each gas sensor corresponds to a specific detected gas, but no gas detector can be absolutely effective. Therefore, when choosing a gas sensor, it is important to understand as much as possible the detection interference of other gases on the sensor to ensure accurate detection of specific gases.

 

C, Pay attention to the lifespan of various portable gas alarm sensors

All types of gas sensors have a certain service life, that is, lifespan. Generally speaking, in portable instruments, LEL sensors have a longer lifespan and can be used for about three years; The lifespan of the photoionization detector is four years or longer; The lifespan of electrochemical specific gas sensors is relatively short, usually between one to two years; The lifespan of oxygen sensors is the shortest, about one year. The lifespan of electrochemical sensors depends on the drying of the electrolyte, so if they are not used for a long time, sealing them in a lower temperature environment can extend their service life to a certain extent. Fixed instruments have a relatively large volume and longer sensor lifespan. Therefore, sensors should be tested at all times and used within their effective period as much as possible. Once they fail, they should be replaced in a timely manner.

 

D, Pay attention to the concentration measurement range of gas detection instruments

The portable gas detector has a fixed detection range for various toxic and harmful gas detectors. Only by completing the measurement within its measuring range can the instrument ensure accurate measurement. However, measuring beyond the measurement range for a long time may cause permanent damage to the sensor. For example, if the LEL detector is accidentally used in an environment with more than 100% LEL, it may completely burn out the sensor. Toxic gas detectors, when used at high concentrations for extended periods of time, can also cause damage. Therefore, if a fixed instrument emits an over limit signal during use, the measurement circuit should be immediately turned off to ensure the safety of the sensor.

 