Preparation before using the noise meter

Oct 11, 2022

Leave a message

Measurement preparation

(1) Loading the battery


Open the battery back cover and insert a 9V battery.


(2) Change the battery


When the voltage of the battery is lower than the operating voltage, the "BAT" symbol will appear. At this point, a new battery should be replaced.


Calibration process

(1) Set to the following keys and function switches.


Display: dB, A, Hi or Lo, F;


Function: A-sound level frequency weighting;


Response time: FAST;


Range level: 30 to 100dB(Lo) or 60 to 130dB(Hi);


Measurement method: Exit the maximum value hold and data hold functions.


(2) Carefully insert the microphone into the acoustic calibrator.


(3) Open the battery cover, take out the battery, and calibrate with a small screwdriver.


The meter has been calibrated before shipment, and the re-calibration cycle is recommended: 1 year.