Repair and Maintenance of Insulation Resistance Multimeters

Nov 22, 2026

Repair and Maintenance of Insulation Resistance Multimeters

 

① The insulation resistance multimeter generally adopts automatic calibration technology. Except for components marked as replaceable, do not replace other electronic components arbitrarily, otherwise the technical specifications may deviate. Below, Guangzhou Yingxiangjia Instrument & Meter Co., Ltd. shares the maintenance procedures for insulation resistance multimeters.

 

1. General Maintenance

① Clean the instrument housing regularly with a damp cloth and neutral cleaner. Do not use abrasives or solvents. Dirt on terminals and probe terminals may affect readings; use cleaner to remove contaminants from each terminal. ② Press the POWER button to switch off the instrument when not in use. Remove the batteries if the instrument will be stored for a long time. ③ Store the instrument away from dampness, high temperature and strong magnetic fields.

 

2. Battery Replacement

① Press the POWER button to turn off the power, disconnect the test probes or test leads from the terminals, and fully separate the test leads from the circuit under test. ② Remove the two screws of the battery cover and take off the battery cover. ③ Take out the batteries from the battery compartment. ④ Install six new AA (AM3 / LR6) batteries (nominal capacity approx. 2450 mAh). ⑤ Refit the battery cover and fasten the two screws.

 

3. Fuse Replacement

To avoid electric shock or arc hazards that may cause personal injury or instrument damage, replace the fuse strictly following the steps below. Only replace it with a fuse of the same model or identical electrical rating.

① Press the POWER button to turn off the power, disconnect the test probes or test leads from the terminals, and fully separate the test leads from the circuit under test. ② Remove the two screws of the battery cover, two screws of the rear housing cover and one screw behind the battery compartment; then remove the battery cover and rear housing cover. ③ Gently pry up one end of the fuse, take out the blown fuse, and install a new fuse of the same model or electrical rating. Ensure the fuse is securely clamped in the middle of the fuse holder. Fuse specification: F1A 240 V. ④ Refit the battery cover and rear housing cover, then tighten the four screws. Do not use the instrument before all screws are fastened.

 