Technical parameters of handheld oscilloscope
1. Human-computer interface
Display color or monochrome 5.7" LCD (115x86mm)-320x240LED backlight
Display mode: 500 real-time sample point display/vector band interpolation
Waveform display: 2 real-time waveforms + 2 reference waveforms/partition display & full-screen display (waveform area up to 110x74)
Screen control touch screen/window menu and graphic commands
Language 5 languages, menu & online help (French, English, German, Italian, Spanish)
2. Oscilloscope mode
vertical
Bandwidth 60MHz Bandwidth limit 15MHz, 1.5MHz or 5kHz
Number of channels: 2 isolation channels
Input impedance 1MΩ±0.5%, 17pF
Maximum input voltage 600V/CatIII (probe connection safety adapter)
Vertical sensitivity: 16 ranges, from 2.5mV to 200V/div. Vertical amplification mode can reach 156μV/div (12-bit converter) accuracy ±1%
Vertical magnification Click window selection magnification system (use stylus to select directly on display) X16 Max
The probe attenuation coefficient is 1/10/100/1000 or adjusted to any value/the measurement unit can be freely defined
3. Level
Time base scans 35 steps, from 1ns/div to 200s/div, accuracy ±0.1% from 100ms to 200s/div in rolling mode
Horizontal magnification Click the window to select the magnification system (use the stylus to select directly on the display)
4. Trigger
Trigger mode automatic, trigger, single, automatic 50%
Trigger type edge, pulse width (20ns-20s), delay (120ns-20s), count (3-16384 events) TV frame or line number (525=NTSC or 625=pAL/SECAM) trigger level continuously adjustable
Coupled AC, DC, high frequency limiting, noise reduction - trigger suppression (from 160ns to 30s)
Sensitivity≤0.6div(1kHz)/≤1.5div(entire bandwidth)
Threshold triggering starts at one of 16 automatic measurements
5. Number processing
Maximum sampling rate equivalent sampling rate 50Gs/s, real-time sampling rate per channel 2.5Gs/s or 1Gs/s
Vertical resolution 12 bits (vertical resolution 0.025%)
Storage depth 2500 points/channel up to 50000 points/channel (optional 50K storage depth expansion)
User-disposable storage space Windows interface file management 2Mb, which can store various files: waveforms, configuration settings, mathematical formulas, picture files, etc.
GLITCH mode pulse width ≥ 2ns, 1250 pairs Min/Max (25000 pairs, when equipped with 50K storage depth expansion)
Display modes Envelope mode, Average mode (coefficients from 2 to 64), XY mode
Streaming, other functions
Automatic triggering time: <5s, automatic identification of input channel - frequency>30Hz
FFT Analysis & Number Operations FFT (linear or logarithmic), measurement pointers/mathematical operations +,-,x,/and custom mathematical formulas
Pointers 2 or 3 pointers: V, T, phase; resolution 12 bits; display: 4 words
Automatic measurement of 19 parameters; resolution 12 bits; display: 4 words
7. Multimeter mode
General indicators 2 channels/4000 words, + bar chart min/max/TRMS/trend chart record with time (5 minutes to 31 days)
AC, DC, AC+DC voltage 300mV to 600Vrms, 400mV to 800VDCVDC accuracy: 0.5%R+5 words, bandwidth 200kHz
Resistance 80Ω~32MΩ, accuracy 0.5%R+25 words, fast continuity test: 10ms
Other measurement capacitance: 5nF to 5mF, frequency: 200kHz, diode test: 3.3V
Measurement threshold triggers 2 channels, time can be set, 100 sets of timestamp records (.TXT file)
8. Harmonic analysis mode (optional)
Multi-channel analysis 2 channels, 61st harmonics, fundamental frequency from 40 to 450Hz (automatic or manual mode)
Simultaneous measurement of total Vrms and THD, harmonics of selected order (% of fundamental, phase, frequency, Vrms)
9. Recording mode (optional)
Record length/sampling rate from 2s to one month, from 800μs to 18min (from 40μs to 53s, when equipped with 50K storage depth expansion)
Recording condition threshold or limit window, multi-channel conditions act at the same time, the period can be set from 160μs
Record analysis scales and physical units, automatic or manual pointer measurements, error search, magnification and more