What are the gas detectors' detection standards?

Dec 01, 2022

Leave a message

What are the gas detectors' detection standards?


1. Appearance and functional inspection


1. Detect appearance and other items


Detect the appearance of the gas detector, which is to avoid minor problems in the transportation or assembly process of the gas detector. We need to check whether the appearance of the gas detector is flawed, cracked or damaged, and check whether the structure of the entire gas detector is whole. At the same time, check the machine model, label, manufacturer name, and delivery time on the gas detector body and check them with the manual or the information given by the manufacturer to ensure accuracy. At the same time, check the explosion-proof mark and measurement license of the gas detector The content such as logo and number must be complete and clear, and some certificates can be provided by the manufacturer.


2. Power-on inspection


The gas detector needs power to work, and it is usually powered by a built-in battery. We need to turn on the switch to check whether the gas detector is powered on normally. Some gas detectors continue to work by replacing the battery, and some gas detectors The detector is equipped with a charger. For the gas detector equipped with a charger, we need to test whether the charger is charged normally. If the power is normal, we need to check whether the display screen of the gas detector is normal.


3. Check whether the sound and light alarm of the instrument is normal


For gas detectors with audible and visual alarm signals, since they are powered by batteries, when undervoltage is displayed, they should be able to send out audible or optical indication signals that are clearly different from the alarm signals.


2. Indication error


The gas detector we bought is used to detect the gas concentration. It is impossible for the gas detector to display the gas concentration perfectly. It has errors, but this error has a range. If it exceeds this range, it means This gas detector does not meet the standard, and the indication error specified by it is different for different gases. For example, it is normal for the indication error of oxygen to be within ±0.5%VOL.


3. Alarm error


We mentioned the error of the displayed value above, so there is a certain allowable error for the alarm value of the gas detector, because the instrument will be affected by various factors, it is impossible to alarm at an accurate concentration every time, Therefore, the concentration of the alarm is allowed to have errors, as long as the errors are within the standard range. Its alarm error is also different for different gases, for example: the alarm error of oxygen is within ±0, 1%VOL.


4. Response time


Response time refers to the time required for the indication value of the gas detector to rise from zero to 90% of the stable indication value that the instrument should reach. This time is also required by the standard. This standard is the same as the indication error and the alarm error. Different gases respond differently.


5. Insulation withstand voltage


For gas detectors, there are still some insulation withstand voltage standards that must be met. Standard requirements: at room temperature: ≥100MΩ; after damp heat: ≥1MΩ. The dielectric strength should be able to withstand 500V AC voltage for lmin, and discharge and breakdown should not occur. Only after these are met can the gas detector meet the standard.