What should be noted when measuring with a pH meter sensor?
Before operation, inspect the integrity of the electrode first. Most electrodes matched with pH meters are combination electrodes. Older-generation pH meters still use glass electrodes and calomel electrodes. Since combination electrodes are widely used, the discussion below mainly focuses on combination electrodes.
Combination electrodes used in laboratories are mainly divided into fully sealed and non-sealed types. Fully sealed types are relatively rare and mostly manufactured by overseas enterprises. Before using a combination electrode, first check whether cracks or breakages exist on the glass bulb. If intact, perform two-point calibration with pH buffer solutions. If both the positioning and slope knobs can be adjusted to the corresponding pH values, the electrode is generally considered usable. Otherwise, activate the electrode in accordance with the operating manual. Activation procedure: Immerse the electrode in 4% hydrogen fluoride solution for about 3–5 seconds, take it out and rinse with distilled water. Then soak it in 0.1 mol/L hydrochloric acid solution for several hours, rinse thoroughly with distilled water, and proceed with calibration. Use the buffer solution with a pH value of 6.86 (25°C) for positioning. After adjustment, select another pH buffer solution arbitrarily for slope adjustment. If adjustment cannot be completed, the electrode needs to be replaced.
For non-sealed combination electrodes, external reference solution (3 mol/L potassium chloride solution) must be filled inside. It is necessary to check whether the potassium chloride solution level is above 1/3 of the electrode cavity. If not, add 3 mol/L potassium chloride solution. If the potassium chloride solution exceeds the small hole, shake off excess solution so that the liquid level is below the hole. Check for air bubbles inside the solution. If bubbles exist, gently tap the electrode to completely expel them.