Operation Methods of Thermal Anemometers
Using the Thermal Anemometer:
① Checking whether the pointer of the meter points to the zero position before use. If any offset exists, gently adjusting the mechanical zero screw of the meter to return the pointer to zero. Setting the calibration switch to the OFF position.
② Inserting the probe rod plug into the socket, placing the probe rod vertically upward, tightening the screw plug to seal the probe, setting the "calibration switch" to the full‑scale position, and slowly adjusting the "full‑scale adjustment" knob until the meter pointer indicates full scale.
③ Setting the "calibration switch" to the "zero" position, and slowly adjusting the "coarse adjustment" and "fine adjustment" knobs until the meter pointer points to zero.
④ After completing the above steps, gently pulling the screw plug to expose the probe of the measuring rod (the exposed length can be selected as required), aligning the red dot on the probe facing the incoming wind, and checking the calibration curve according to the meter reading to obtain the measured wind speed.
⑤ Upon finishing measurement, setting the "calibration switch" to the OFF position.
The above covers the operating procedures of the thermal anemometer. The following section further introduces its applications.
Applications of Thermal Anemometers
① Measuring the velocity and direction of mean flow.
② Measuring fluctuating incoming velocity and its frequency spectrum.
③ Measuring Reynolds stress in turbulent flow, as well as two‑point velocity correlation and temporal correlation.
④ Measuring wall shear stress (usually performed by using a hot‑film probe flush‑mounted on the wall, working on a principle similar to hot‑wire velocimetry).
⑤ Measuring fluid temperature (calibrating the resistance‑temperature curve of the probe in advance; the fluid temperature can be determined from the measured probe resistance).
An anemometer is an instrument for measuring air flow velocity. There are many types. Cup anemometers are widely used at meteorological stations. Their sensing assembly consists of three conical cups mounted at 120° intervals on a support frame, with the concave side of each cup facing the same direction. The entire sensing unit is installed on a vertical rotating shaft. Driven by wind force, the cups rotate around the shaft at a rotational speed proportional to the wind speed.