Principle Analysis of Current Measurement by Clamp Meters
A clamp meter essentially consists of a current transformer, a clamp trigger, and a rectified moving‑iron repulsion instrument.
The operating principle of a clamp meter is the same as that of a transformer. The primary winding is the conductor passing through the clamp core, equivalent to a one‑turn primary coil of a step‑up transformer. The secondary winding and the measuring ammeter form the secondary circuit. When alternating current flows through the conductor, this single‑turn coil generates an alternating magnetic field, which induces current in the secondary circuit. The ratio of the induced current to the primary current is inversely proportional to the turns ratio of the primary and secondary windings. Clamp ammeters are designed for measuring large currents. If the current to be measured is small, the primary conductor can be wrapped several times through the clamp meter; divide the measured reading by the number of turns to obtain the actual current.
The secondary winding of the through‑core current transformer inside the clamp ammeter is wound on the iron core and connected to the AC ammeter. Its primary winding is the test conductor passing through the center of the transformer. The rotary knob acts as a range selector switch. The trigger opens and closes the movable part of the through‑core transformer core to clamp onto the test conductor.
For current measurement, squeeze the trigger to open the jaws and place the live test conductor in the center of the through‑core current transformer. When alternating current flows in the conductor, the magnetic flux of the alternating current induces a current in the secondary winding of the transformer. This induced current passes through the coil of the electromagnetic ammeter, deflecting the pointer to indicate the measured current value on the dial scale.
After placing the test conductor into the window by operating the core trigger, ensure the two jaw surfaces fit tightly with no foreign objects in between. The minimum range of this clamp meter is 5 A, which will produce relatively large errors when measuring small currents. In this case, wrap the live conductor around the clamp meter several times before measurement. Divide the obtained reading by the number of turns to get the true value.