Troubleshooting Methods for Common Faults of Online pH Meters

Oct 04, 2026

Troubleshooting Methods for Common Faults of Online pH Meters

 

1. Aging of Glass Electrode

Phenomena: Slow response; reduced sensitivity; zero drift.

Treatment: (1) Perform electrode calibration regularly. (2) Electrode activation: Immerse the aged electrode in a mixed solution of 1M acetic acid and 1M potassium chloride (1:1) for 10 minutes, then take it out and rinse thoroughly. For electrodes with slight aging, soak them in distilled water or 0.1N dilute hydrochloric acid solution for 24 hours for activation.

 

2. Contamination of Glass Electrode

Phenomena: Reduced sensitivity; measurement deviation.

Treatment: (1) Alkaline precipitates on the electrode can be removed by cleaning in dilute hydrochloric acid solution, then rinse the electrode thoroughly with distilled water. (2) Grease attached to the electrode can be cleaned with hot water and household detergent, or wiped with gauze. (3) Colloidal contaminants can be cleaned with strong hydrochloric acid solution. Rinse the electrode completely with distilled water after cleaning.

After cleaning with acidic, alkaline or organic solvents, soak the electrode in distilled water for a period of time to restore the damaged hydrated layer.

 

3. Contamination of Reference Electrode (Ceramic Junction Blockage)

Phenomena: High measured readings; unstable indication.

Treatment: Clean with hot water and household detergent. Oily or alkaline contaminants can be removed with organic solvents (e.g. alcohol) or dilute hydrochloric acid solution. For heavy contamination, scrub gently with a soft brush or wipe with filter paper strips. For non-flowing electrodes, place the electrode in electrolyte at 80°C until the electrolyte cools down.

 