Troubleshooting of Common Faults for Industrial pH Controllers

Oct 04, 2026

Troubleshooting of Common Faults for Industrial pH Controllers

 

① Moisture in the junction box of industrial pH controller: Open the junction box cover of the industrial pH controller. Check whether there is accumulated water or moisture inside the box. Ensure the box is dry and verify that the sealing performance of all parts of the junction box is intact, then perform calibration. If calibration succeeds, the fault is resolved; if not, refer to Step ②.

 

② Incorrect stripping of electrode cable joints of industrial pH controller: Open the rear cover of the industrial pH controller main unit and the junction box cover. Check if the stripping of the pH electrode cable is correct. A common issue is incomplete removal of the black conductive rubber from the cable. There are three positions in total: two inside the junction box and one inside the rear cover of the main unit. Make sure the rubber is fully stripped, rewire and carry out calibration. If calibration succeeds, the fault is resolved; if not, refer to Step ③.

 

③ Water ingress in electrode protective sleeve: Remove the pH electrode extension cable. Move the pH electrode together with its entire protective sleeve to the main unit. Directly connect the core wire and mesh wire of the pH electrode to the GLASS and REF terminals of the industrial pH controller main unit respectively, then perform calibration. If calibration succeeds, refer to Step ⑤. If not, cut off the exposed section of the pH electrode cable inside the junction box, re-strip the cable as required in Step ②, reconnect to the main unit and calibrate. If calibration still fails, the pH electrode is damaged by moisture. Open the pH electrode protective sleeve for inspection. If water enters the sleeve, moisture occurs or the pH electrode is broken, replace the pH electrode.

 

④ Damaged pH electrode of industrial pH controller: Check whether the pH electrode is broken. Replace the pH electrode if damaged.

 

⑤ Damaged and waterlogged extension cable of industrial pH controller: If calibration succeeds after rework, cut off the exposed section of the extension cable inside the junction box, re-strip the cable following the requirements in Step ②, connect and calibrate. If calibration fails, the extension cable needs to be replaced.

 