Application Methods of pH Meters in Dairy Testing

Oct 04, 2026

Application Methods of pH Meters in Dairy Testing

 

Methods for cleaning the pH meter after measuring dairy products, including cleaning for ordinary protein and high-protein samples. Different media are measured in the testing process, so the cleaning method shall be analyzed according to specific conditions. There are several categories of cleaning procedures for pH meters to remove deposits left on the electrode.

 

① When measuring ordinary dairy proteins: Soak the electrode in pepsin hydrochloric acid solution (HCl concentration: 0.1 mol/L) for 15 minutes, or soak it in 10% sodium hypochlorite solution for 5 minutes, then soak in the soaking solution for 12 hours to balance the electrode potential. This can restore the original activity of the electrode. If the effect is unsatisfactory, further inspection is required.

 

② When measuring high-protein and high-fat dairy products: Soak the electrode in dishwashing detergent for 1 hour. Take out the electrode and rinse thoroughly, then immerse it in pepsin solution mixed with 0.5 mol/L hydrochloric acid solution at a volume ratio of 1:1. After treatment, take out the electrode, rinse thoroughly with distilled water, and then soak it in 3 mol/L solution. Do not allow the electrode to dry out.

Dairy products feature high protein and high fat content, which can easily clog the capillary tube of the electrode glass bulb. This is a common phenomenon during normal operation. It helps save operating costs and allocate resources reasonably.

 

pH meters are widely applied in the dairy industry. They play a vital role in the testing of semi-finished and finished products. Failure to obtain accurate pH values timely for production reference will cause immeasurable impacts on production.

 