What are the technical requirements for Clamp Meter Calibration Device?

May 26, 2023

Leave a message

What are the technical requirements for "Clamp Meter Calibration Device"?

 

AC output


Voltage range 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V;


Current range 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A;


Adjustment range 0~120% (except 1000V voltage);


Voltage and current output capacity up to 20VA;


Voltage and current adjustment accuracy


A: Voltage 20~1000V, current 0.2~20A range


±(0.03%RD+0.02%RG);


B: voltage 200 ~ 4000mV, current 2 ~ 40mA range


±(0.3%RD+0.2%RG);


DC output


Voltage range 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V,


200V, 400V, 1000V;


Current range 20μA, 40μA, 200μA, 400μA, 2mA, 4mA, 20mA, 40mA,


0.2A, 1A, 2.5A, 5A, 10A, 20A;


Voltage and current adjustment range 0~110% (except 1000V voltage);


Voltage and current output capacity up to 20W;


Voltage adjustment accuracy ±(0.03%RD + 0.02%RG);


Current setting accuracy ±(0.06%RD + 0.04%RG);


Point frequency square wave


Point frequency square wave output 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz;


Square wave amplitude 4.5V.


Electronic resistance box output


·Measuring range 0~200Ω, 0~2kΩ, 0~20kΩ, 0~200kΩ, 0~2MΩ;


·Adjust fineness 0.025%;


·Setting accuracy ±(0.12%RD + 0.08%RG)

 