Brief introduction to the requirements of carbon monoxide gas detector

 

1. The measurement range of the carbon monoxide gas detector should reach 0-2000umol/mol, and the maximum measurement value should not be less than 2000umol/mol.

2. The error of the indication value of the carbon monoxide gas detector should be within the range of ±5umol/mol, and the relative error should be within the range of ±10%. Accurate readings can better indicate the concentration of the gas.

3. The response time of the carbon monoxide gas detector should meet the following standards: the response time of the diffusion carbon monoxide gas detector should be less than or equal to 60 seconds, and the response time of the inhalation carbon monoxide gas detector should be less than or equal to 30 seconds.

4. The drift degree of the carbon monoxide gas detector should be within the following ranges: the zero point drift should be ±3umol/mol; the range drift should be ±5%.

 