How to use the pH meter and maintain the electrode correctly?
1. When the composite electrode is not in use, it can be fully soaked in 3M potassium chloride solution. Do not soak with washing liquid or other water-absorbing agents.
2. Before use, check the bulb at the front end of the glass electrode. Under normal circumstances, the electrode should be transparent without cracks; the bulb should be filled with solution without air bubbles.
3. When measuring a solution with a high concentration, shorten the measurement time as much as possible, and wash it carefully after use to prevent the measured liquid from adhering to the electrode and polluting the electrode.
4. After cleaning the electrode, do not wipe the glass membrane with filter paper, but use filter paper to dry it, so as to avoid damage to the glass membrane, prevent cross-contamination, and affect the measurement accuracy.
5. During the measurement, pay attention to the silver-silver chloride internal reference electrode of the electrode should be immersed in the chloride buffer solution in the bulb to avoid the digital jumping phenomenon in the display part of the electric meter. When using, pay attention to gently shake the electrode a few times.
6. The electrode cannot be used in strong acid, strong alkali or other corrosive solutions.
7. It is strictly forbidden to use in dehydrating medium such as absolute ethanol, potassium dichromate, etc.